Thesis Defence: Development of an Electrified Penile Implant with Wireless Control for Enhanced User Experience
October 23 at 11:00 am - 3:00 pm

Parham Fatemi, supervised by Dr. Hadi Mohammadi, will defend their thesis titled “Development of an Electrified Penile Implant with Wireless Control for Enhanced User Experience” in partial fulfillment of the requirements for the degree of Master of Applied Science in Mechanical Engineering.
An abstract for Parham Fatemi’s thesis is included below.
Defences are open to all members of the campus community as well as the general public. Please email hadi.mohammadi@ubc.ca to receive the Zoom link for this defence.
Abstract
Erectile dysfunction (ED) is a widespread condition affecting millions of men worldwide, with its prevalence expected to increase in the coming years. Although conventional penile prostheses are clinically effective, their fundamental design and operating principles have changed relatively little over the past several decades. As a result, existing devices may still present limitations related to natural feel, mechanical reliability, ease of operation, and overall user experience.
This thesis investigates the integration of soft robotics and electrical engineering technologies, particularly micropumps, electronically controlled fluidic systems, and wireless actuation, into the development of a next-generation penile prosthesis. The primary objective is to enhance both device functionality and user interaction through an interdisciplinary approach combining soft robotic actuation, embedded electronics, fluidic control, and biomimetic structural design.
The study evaluates the feasibility and performance of soft robotic components designed to provide both axial and radial expansion. Multiple prototypes were developed using different materials and structural configurations, including porous architectures intended to reduce the required fluid volume while maintaining sufficient expansion capacity. An electrically controlled pressurization system was also incorporated, with force-sensitive resistors used for calibration and feedback. Wireless control through Bluetooth communication enables remote and discreet operation using a smartphone interface.
Bench-scale testing demonstrated that the prototype was capable of reaching a target pressure of approximately 100 mmHg within 30 seconds following a wireless activation command. This electronically controlled approach eliminates the need for the repeated manual pumping associated with conventional inflatable penile prostheses and demonstrates the potential for a more automated and user-friendly operating system. In addition, a specially designed external shell intended to mimic the mechanical function of the tunica albuginea was incorporated to improve the structural response, realism, and biomimetic behavior of the device.
Despite several remaining engineering challenges, including the need for improved thermal management of the solenoid valves and further optimization and miniaturization of system components, the results demonstrate the potential of combining soft robotics with electronically controlled fluidic systems to advance penile prosthesis technology.
This work represents a bench-scale engineering feasibility study rather than a clinical or preclinical evaluation. No human or animal subjects were involved, and the current prototype incorporates materials and hardware that are not intended for implantation. Within these limitations, the study establishes and validates a functional actuation and control architecture that may serve as a foundation for the future development of implantable devices in this class.